Search our store

£207.80 £249.36 (Inc. VAT)

Product Description

Pocket Waterproof EC, pH and TDS (Low Range) tester

Multiparameter pH, EC, TDS and °C Tester

Part no: HI 98129 (low range)

Part no: HI 98130 (high range)

HI 98129 and HI 98130 are waterproof testers that offer
high accuracy pH, conductivity, total dissolved solids
and temperature measurements in a single meter.

No more switching between meters for routine measurements:

  • These floating, waterproof combination testers have an easy to read display and automatic shut-off
  • pH and EC/TDS readings are automatically temperature compensated
  • The EC/TDS conversion factor is user selectable as well as the temperature compensation coefficient
  • Both testers feature a replaceable pH electrode cartridge together with an EC/TDS graphite electrode
  • Automatic shut-off helps preserve battery life
  • Replaceable pH cartridge extends life of the tester – when pH sensor is exhausted, simply replace it with a new one
  • Calibrate direct into Hanna buffer solution sachets
  • Automatic 1 or 2 point pH calibration
  • Rubberised keyboard
  • Quick and easy to operate
  • Easy switch between Celsius & Fahrenheit readouts
  • HOLD function
  • Easy to replace pH electrode cartridge with sturdy snap-in connector means there are no pins to bend or break
  • Large LCD display
  • pH and temperature displayed simultaneously
  • Automatic temperature compensation
  • BEPS Battery Error Prevention System
  • Floating waterproof casing with tactile grip makes testing easier and less of a headache
  • Removable safety cap

Main benefits:

  • On-screen battery life
  • LCD indicates the percentage of battery power remaining at startup
  • Instability & ATC indicators
  • Ensures reliable EC and TDS measurements. ATC symbol is shown when active
  • Adjustable temperature coefficient factor
  • Users can choose between different BETA (B) factors for precise temperature compensated measurements
  • HOLD function
  • The HOLD function freezes the LCD display temporarily
  • Standard or N.I.S.T buffer calibration
  • Automatic calibration is performed with 2 sets of memorised buffers for greater accuracy
  • Adjustable TDS ratio
  • For measurement accuracy, users can choose between different conductivity to TDS conversion factors

Low or High range conductivity?
Simply select the model that is most appropriate for your testing range:

  • For low range: select HI 98129
  • 0 to 3999 uS/cm; 0 to 2000 ppm
  • For high range: select HI 98130
  • 0.00 to 20.00 mS/cm; 0.00 to 10.00 ppt

Supplied with:

  • HI 73127 pH electrode
  • HI 73128 electrode removal tool
  • Batteries
  • <

  • Instructions

High accuracy EC/TDS graphite probe

The graphite conductivity probe provides greater accuracy because it cannot be contaminated by salt deposits in the solution.
The exposed temperature sensor provides fast response times and guarantees highly accurate temperature compensated readings

Technical Data
Range pH 0.00 to 14.00pH
Range EC 0 to 3999 µS/cm
Range TDS 0 to 2000 ppm
Range Temperature 0.0 to 60.0°C / 23.0 to 140.0°F
Resolution pH 0.01 pH
Resolution EC 1 µS/cm
Resolution TDS 1 ppm
Resolution Temperature 0.1°C / 0.1°F
Accuracy pH pH±0.05 pH
Accuracy EC/TDS ±2% F.S.
Accuracy Temperature ±0.5°C / ±1°F
Calibration pH Automatic at 1 or 2 points with 2 sets of memorised pH buffers
(pH 4.01 / 7.01 / 10.01 or 4.01 / 6.86 / 9.18)
Calibration EC/TDS automatic, 1 point
Temperature Compensation pH: automatic; EC/TDS: automatic with ß adjustable from 0.0 to 2.4% / °C
TDS Conversion Factor adjustable from 0.45 to 1.00
pH Electrode HI 73127 (replaceable; included)
Battery Type/Life 4 x 1.5volt giving approx 100 hours continuous use
, auto-off after 8 minutes of non-use
Environment 0 to 50°C (32 to 122°F); RH max 100%
Dimensions 163 x 40 x 26mm,
Weight 100 g

td

Related Products