£76.91 £92.29 (Inc. VAT)
LogTag TRIL-8 Low Temperature Data Logger
TRIL-8 Low Temperature “probe less”
Multiple Use Recorder
Record temperatures in dry ice shipments or for
ULT freezer validation from -80°C to +40°C,
without the need of external sensors.
1000 hour operation in dryice at -80°C
The LogTag TRIL-8 Dry Ice €œProbe-less€ Temperature Recorder operates, measures and stores up to 8000 temperature readings in temperature environments ranging from -80°C to +40°C (-112°F to +104°F).
Intended for use in transit monitoring of articles stored in packaging incorporating dry ice cooling agents.
Now supported with an enhanced battery management system that rates the product €˜multiuse€™ for accumulative exposure of up to 1000 hours @ -80°C.
All new TRIL-8 product will require the LogTag Analyzer software version 2.2r20 or later.
Using the LogTag Interface and LogTag€™s freely available companion software LogTag Analyzer, the LogTag Dry Ice Temperature Recorder is easily configured for recording conditions including delayed start, sampling interval, number of readings and configuration of conditions to activate the ALERT indicator.
The Red Alert indicator provides an immediate indication, without access to a PC, if any readings are outside the limits specified at the time the unit was configured.
Green OK indicator provides immediate visual confirmation, without access to a PC, that the unit is operating.
Readings are downloaded using LogTag Analyzer, which provides facilities for charting, zooming, listing data statistics and allows exporting the data to other applications such as MS Excel.
The product incorporates an external temperature sensor uniquely arranged in a recess in the case. This arrangement protects the sensor from physical damage while retaining the faster response time normally associated with remote sensors.
Features
To export the data from your LogTag to the LogTag Analyser software, a LogTag USB Docking Station / Interface Cradle is required.
Technical Data | |||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Model | TRIL-8 (New version) – 1000 hours continuous use! | ||||||||||||||||||||
Measurement Range | -80 ~ +40°C (-112°F ~ +104°F) | ||||||||||||||||||||
Resolution |
< 0.1°C for -80°C ~ 0.0°C < 0.2°C for 0.0°C ~+20°C < 0.5°C for +20°C ~+40°C |
||||||||||||||||||||
Accuracy | Better than ±1°C for -30°C~ +20°C better than ±1.2°C for -45°C~ -30°C & +20°C~+40°C better than ±1.7°C for -80°C~–40°C |
||||||||||||||||||||
Sensor reaction time | typically less than 5 minutes(T90) in moving air (1m/s). | ||||||||||||||||||||
Capacity | 8000 temperature readings (16Kbytes memory) | ||||||||||||||||||||
Sampling frequency | adjustable, 1~60 minutes or 1~18 hours | ||||||||||||||||||||
Logging start options | Push button start or specific date & time. | ||||||||||||||||||||
Recording indication | Flashing ‘OK’ indicator flashing ‘Alert’ indicator. | ||||||||||||||||||||
Download Time | Typically with full memory (8000 readings) in less than 5 secs depending on computer or readout device used. | ||||||||||||||||||||
Environmental | IP65 (roughly equivalent to NEMA 4) | ||||||||||||||||||||
Power source | 3V low temperature chemistry Lithium battery. | ||||||||||||||||||||
Battery life | Minimum storage life of 12 months before €˜start€™. Limited use version rated for up to 3 trips of up to 2 weeks duration at dry ice temperatures over a 2 year period. |
||||||||||||||||||||
Size | 86mm(H)x54.5mm(W)x8.6mm(T) | ||||||||||||||||||||
Weight | 35g | ||||||||||||||||||||
Case Material | Polycarbonate |
Sinclair and Campbell is a leading supplier of Digital Temperature Measurement and Monitoring Instruments to the Food Processing and Catering industries.
© 2022 Sinclair Campbell | Hosted & Managed by RocketWP
Registered Office: SINCLAIR & CAMPBELL,
CLACHAN HOUSE, ROSNEATH, HELENSBURGH, SCOTLAND UK, G84 0QR